CZ  |  EN  |  SK

Water purification systems (Grade 1) for bio applications

Adrona

Conductivity (Grade 1): <0,055 µS/cm
TOC: <2 ppb
DNase: <4 pg/µl
RNase: <0,01 ng/ml
Bacteria: <1 CFU/ml
Endotoxins: <0,001 EU/ml
Particles (>0,22 µm): <1 per ml

The Crystal EX Bio and Integrity+ Bio systems produce ultrapure water right from tap water. Water is purified in two stages: pure water is produced first and stored in the tank. Pure water is later used for the production of ultrapure water, used for highly sensitive biological and chemical applications like flow cytometry, cell and tissue culture and molecular biology.

System OneSite+ Bio uses pre-treated water, Integrity+ Bio is for labs with low consumption, E30 Biosystem has a color display. Q-Front Bio is a new modern system with low running costs. 

Grade 1 water produced by these systems complies with ISO 3696, ASTM and CLSI standards.

For certain models, other Trace, HPLC, Bio versions of the system are available.

Contact us